PlastCare USA dental self-sealing sterilization pouches are an essential tool for any dental practice looking to maintain a clean and hygienic environment during procedures. These innovative pouches provide reliable protection against contamination and damage, ensuring a safe and efficient dental experience for patients.

Made from high-quality and durable materials, our dental self-sealing sterilization pouches are designed to fit securely over dental instruments, providing a barrier between the instruments and the patient’s mouth. Their efficient design ensures optimal protection against contamination and damage during procedures.

Our dental self-sealing sterilization pouches are also easy to use, with a simple and reliable application process that ensures a secure fit. They are compatible with most sterilization methods, making them a versatile and convenient addition to any dental practice.

Item # PG-3565

Available Sizes

Specifications

  • After instruments have been cleaned, they must be placed in sterile bags or pouches to ensure that they remain sterile until their next use. This is true for nail technicians, dentists, and other medical professionals. Knowing you’re getting clean and sterile instruments is a very calming thought when sitting in the dentist chair.
  • These self-sterilization pouches help ensure the safety of staff members and dental patients alike. These pouches are approved by the FDA as a medical device to sterilize equipment and protect tools against contamination during handling. The pouches protect the contents from spreading infection or being infected.
  • Not only do these pouches help with sterilization but they also improve organization. These pouches act as organizers as they keep dental tool kits together so all the instruments are in the same place.

Do you know how to load?

Having an autoclave in good working condition is just the beginning. You also need to ensure your teammates know how to load instruments and packs correctly for optimal results.

  • An autoclave should never be overloaded. It is better to run two cycles with smaller loads, than a single cycle with an autoclave filled to the brim. Sadly, overloading happens much more frequently than practitioners realize.
  • Pouches should not be stacked horizontally. Instead, they should be placed flat and sterilized one at a time. This is called “single height loading.”
  • If you place a pouch flat, the paper side should face up, and the plastic side should face down. This allows moisture to escape through the paper side (hot air goes up).
  • Pouches placed vertically (i.e. side by side on a rack) should have enough air between them to allow for ventilation. You can purchase commercial racks made for this purpose.
  • Pouches should not touch the inner sides of the autoclave, as the inside gets hot and can burn them.
  • Ensure instruments are clean and dry before wrapping them.
  • At the end of a cycle, the door should be open as soon as possible to allow the equipment to dry and to prevent items from burning due to excessive heat exposure.
  • Instrument packs should not touch each other in the autoclave so that steam can circulate between them.
  • Remember that sterilization tape only means the surface of a pack was sterilized. It does not ensure the deepest part of the pack is sterile. This is the reason why a sterilization strip must be placed deep inside a pack (not in the top layer, as is often done).

Following these simple guidelines will help ensure true sterilization has been accomplished.